Reproducibility in the Hematology Laboratory: The Microhematocrit Determination 论文
1971American Journal of Clinical Pathology引用 522
Clinical Laboratory Practices and Quality ControlDigital Imaging for Blood DiseasesBlood transfusion and management
详细信息
- 发表期刊/会议
- American Journal of Clinical Pathology
- 发表日期
- 1971-07-01
- 发表年份
- 1971
关键词
Clinical Laboratory Practices and Quality ControlDigital Imaging for Blood DiseasesBlood transfusion and management
摘要
To estimate the overall microhematocrit test error and to determine the relative contributions of reading error, capillary tubes, centrifugation times, and clays to the variation in test results, a prepared erythrocyte suspension was tested repeatedly on six alternate days. Results indicate that the variation caused by reading the hematocrit to the nearest one-half percentage point exceeds that attributable to differences among capillary tubes and centrifugation times.
相关技术
暂无数据
相关事件
暂无数据
相关文章
暂无数据