Reproducibility in the Hematology Laboratory: The Microhematocrit Determination 论文

1971American Journal of Clinical Pathology引用 522
Clinical Laboratory Practices and Quality ControlDigital Imaging for Blood DiseasesBlood transfusion and management

详细信息

发表期刊/会议
American Journal of Clinical Pathology
发表日期
1971-07-01
发表年份
1971

关键词

Clinical Laboratory Practices and Quality ControlDigital Imaging for Blood DiseasesBlood transfusion and management

摘要

To estimate the overall microhematocrit test error and to determine the relative contributions of reading error, capillary tubes, centrifugation times, and clays to the variation in test results, a prepared erythrocyte suspension was tested repeatedly on six alternate days. Results indicate that the variation caused by reading the hematocrit to the nearest one-half percentage point exceeds that attributable to differences among capillary tubes and centrifugation times.